High-performance liquid chromatographic determination of the silicon phthalocyanine Pc 4 in human blood
M M Zuk1, M E Kenney, B Horowitz
1New York Blood Center, NY 10021, USA.
Abstract:
A sensitive and reproducible method has been developed for the measurement of the silicon phthalocyanine Pc 4 in red blood cell concentrates (RBCC). The procedure involves extraction of the RBCC with acetonitrile, purification of the extracts with reversed-phase Sep-Pak C18 cartridges and determination of Pc 4 in the extracts by high-performance liquid chromatography (HPLC) using a reversed-phase C18 column. The detection limit with 1-ml RBCC samples is 2 ng. This method is applicable to monitoring Pc 4 during its use as a photosensitizer for the inactivation of viruses in RBCC prior to transfusion. It has the potential to be adapted for measuring Pc 4 in tissues during its use in photodynamic therapy of cancer.