Related Experiment Videos

The Harris-Galante cementless femoral component: poor results in 57 hips followed for 3 years

F de Nies1, M W Fidler

  • 1Department of Orthopaedics, Onze Lieve Vrouwe Gasthuis, Amsterdam, The Netherlands.

Insights

The Harris-Galante femoral prosthesis showed poor early clinical results, with significant mid-thigh pain and femoral component loosening due to osteolysis in a 3.4-year follow-up study.

Area of Science:

  • Orthopedic Surgery
  • Biomedical Engineering
  • Materials Science

Background:

  • The Harris-Galante prosthesis was used in hip replacement surgeries.
  • Early clinical outcomes and potential complications require thorough evaluation.

Purpose of the Study:

  • To assess the early clinical results and radiographic outcomes of the Harris-Galante femoral prosthesis.

Main Methods:

  • A retrospective review of 68 Harris-Galante femoral prostheses implanted between 1986 and 1991.
  • Follow-up of 57 hips in 53 patients after a mean of 3.4 years.
  • Evaluation of clinical outcomes using the Harris Hip Score (HHS) and radiographic assessment for loosening and osteolysis.

Main Results:

  • The average HHS improved from 39 preoperatively to 81 postoperatively.
  • Only 34 hips (59.6%) achieved excellent or good results.
  • Mid-thigh pain was reported in 30 cases (52.6%).
  • Radiographic loosening of the femoral component occurred in 10 hips (17.5%) due to osteolysis.
  • Seven acetabular cups (12.3%) also showed signs of radiographic loosening.

Conclusions:

  • The Harris-Galante femoral prosthesis demonstrated suboptimal early clinical results, characterized by high rates of mid-thigh pain and femoral component osteolysis.
  • Radiographic loosening of the femoral component was a significant issue.
  • The findings suggest potential concerns regarding the long-term performance and design of this prosthesis.