Artifacts and thresholding in X-ray CT of a cortical bone and titanium composite
Related Experiment Video
Updated: Nov 5, 2026

A Sectioning, Coring, and Image Processing Guide for High-Throughput Cortical Bone Sample Procurement and Analysis for Synchrotron Micro-CT
Published on: June 12, 2020
1Department of Orthopedic Surgery, Washington University, St. Louis, MO 63110, USA.
Purpose:
X-Ray CT has the potential to provide precise and accurate data from which the mechanical properties of bone can be calculated. Such data would be useful in understanding the response of bone tissue to implants. Various artifacts can, however, degrade accuracy of the data. The aim of this study was to measure the artifactual errors produced in CT of a simulated femoral bone-titanium composite and to propose a method to correct for them.
Method:
A composite phantom that simulates cortical bone and a titanium implant was designed and constructed. The phantom was scanned and the image data were analyzed over a range of thresholds with image analysis software developed for this study. The outer (OD) and inner (ID) diameter and the CT number of the cortical bone, with and without titanium, were measured over a range of cortical thicknesses.
Results:
While ID can be accurately measured by choosing the proper threshold (800 HU), OD, even at optimal threshold, will be underestimated by approximately 2%. If a proper threshold is selected, CT number can also be accurately determined. Errors of up to 5%, however, are produced by titanium unless corrected by proper threshold selection.
Conclusion:
Intramedullary titanium is not a deterrent to obtaining accurate measurements of cortical bone dimensions and properties. Proper choice of thresholds for image analysis of CT scan data can yield accuracy and precision of < or = 2%.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...