A calibration program for Rotorod samplers

D A Frenz1, J C Elander

  • 1Multidata, Inc., Minnetonka, Minnesota 55305-1533, USA.

Annals of Allergy, Asthma & Immunology : Official Publication of the American College of Allergy, Asthma, & Immunology
|March 1, 1996
PubMed
Summary

Regular calibration of Rotorod Samplers is crucial for accurate allergy testing. Many devices deviate from manufacturer specifications, impacting data quality and patient care.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Sonic hedgehog regulates otic capsule chondrogenesis and inner ear development in the mouse embryo.

Developmental biology·2002
Same author

A rapid, reproducible method for coating Rotorod Sampler collector rods with silicone grease.

Annals of allergy, asthma & immunology : official publication of the American College of Allergy, Asthma, & Immunology·2001
Same author

Round window membrane delivery of L-methionine provides protection from cisplatin ototoxicity without compromising chemotherapeutic efficacy.

Neurotoxicology·2001
Same author

Growth factor expression in cartilage wound healing: temporal and spatial immunolocalization in a rabbit auricular cartilage wound model.

Osteoarthritis and cartilage·2001
Same author

Growth factor treatment enhances vestibular hair cell renewal and results in improved vestibular function.

Proceedings of the National Academy of Sciences of the United States of America·2001
Same author

Interpreting atmospheric pollen counts for use in clinical allergy: allergic symptomology.

Annals of allergy, asthma & immunology : official publication of the American College of Allergy, Asthma, & Immunology·2001

Area of Science:

  • Environmental Health
  • Allergy Diagnostics
  • Air Sampling Technology

Background:

  • Rotorod Samplers are widely used in the United States for allergy testing.
  • The necessity of regular calibration for these samplers has not been previously established.

Purpose of the Study:

  • To assess the operational conformity of field-used Rotorod Samplers with manufacturer specifications.
  • To investigate the impact of regular calibration on device performance.

Main Methods:

  • Allergists voluntarily submitted their Rotorod devices for calibration over two winter seasons.
  • Motor speed of each sampler was precisely measured using a stroboscopic tachometer.

Main Results:

  • Out of 89 examined Rotorod Samplers, only 10 operated at the exact 2400 RPM specification.
  • An additional 31 devices were within a 5% tolerance range.
  • The majority of devices showed deviations exceeding 5% or had non-operational motors.

Conclusions:

  • The study highlights the critical need for regular calibration of Rotorod Samplers.
  • Maintaining manufacturer specifications is essential for ensuring reliable allergy data.
  • Calibration is vital for consistent and high-quality diagnostic results.