Randomized comparison of ultrasonic aspiration versus conventional electrocautery for dissection of the human

K L Blake1, P A Watt, J M Smith

  • 1Department of Medicine, University of Leicester, United Kingdom.

Insights

Ultrasonic aspirator use during internal thoracic artery harvesting for coronary artery bypass grafting shows reduced vessel trauma compared to conventional electrocautery. This method preserves arterial function and improves responses to certain vasodilators.

Area of Science:

  • Cardiovascular Surgery
  • Vascular Biology
  • Surgical Techniques

Background:

  • Conventional electrocautery is the standard for internal thoracic artery (ITA) harvesting in coronary artery bypass grafting (CABG).
  • Potential for vessel damage exists with conventional electrocautery, impacting graft patency and function.
  • Alternative harvesting techniques are being explored to minimize ITA trauma.

Purpose of the Study:

  • To compare the efficacy and safety of ultrasonic aspirator-assisted ITA harvesting versus conventional electrocautery.
  • To evaluate the functional integrity of ITA segments harvested using different methods.
  • To assess the impact of harvesting techniques on vascular reactivity.

Main Methods:

  • Patients were randomized into six groups: conventional electrocautery, ultrasonic aspirator (60%, 80%, 100% power), and ultrasonic aspirator (100% CAVI-Pulse modes 1 and 3).
  • Harvested ITA ring segments were subjected to organ bath studies to assess contractile responses (potassium chloride, noradrenaline) and relaxation responses (acetylcholine, bradykinin, sodium nitroprusside).

Main Results:

  • Conventional electrocautery resulted in 40% of vessels failing to respond to contractile stimuli due to trauma.
  • Ultrasonic aspirator groups showed significantly reduced failure rates (10% at 60-100% power, 0% in CAVI-Pulse mode 1).
  • Ultrasonic aspirator use enhanced contractile responses and improved bradykinin-mediated relaxation compared to conventional methods.

Conclusions:

  • Ultrasonic aspirator-assisted ITA harvesting, especially at 100% power and CAVI-Pulse mode 1, minimizes vessel damage and preserves vascular function.
  • This technique offers a promising alternative to conventional electrocautery for CABG, potentially improving graft outcomes.
  • Further clinical studies are warranted to confirm the long-term benefits of this improved harvesting technique.