In vivo validation of intravascular ultrasound length measurements using a motorized transducer pullback system
R T Fuessl1, G S Mintz, A D Pichard
1Intravascular Ultrasound Imaging and Cardiac Catheterization Laboratories, Washington Hospital Center, Washington, DC, USA.
The American Journal of Cardiology
|May 15, 1996
Abstract:
Using sonoreflective endovascular targets of known length (stainless steel tubular slotted stents), we have validated in vivo the accuracy and reproducibility of intravascular ultrasound length measurements using a system incorporating motorized transducer pullback through a stationary imaging sheath. The correlation was r = 0.936, with a measurement error of only +/- 5.2%, minimal intraobserver variability, and variability of sequential measurements of only +/- 4.8%.