Dialyzer reuse: interaction between dialyzer membrane, disinfectant (formalin), and blood during dialyzer reprocessing
Related Experiment Videos
1Department of Medicine, Veterans General Hospital, Taipei, Taiwan, Republic of China.
Abstract:
The growing practice of dialyzer reuse in recent years is mainly based on medical and economic considerations. However, adverse reactions such as immunohemolytic anemia due to anti-Nform antibody associated with dialyzer reuse have been reported. In this study, scanning electron microscopy and cytologic staining were used to evaluate the interaction between blood components and the reprocessed synthetic dialyzer membrane (polysulfone) after disinfectant (formaldehyde) treatment. The results showed that various blood components such as fibrin and blood cells still adhered to the dialyzer membrane after reprocessing. The study also demonstrated that the adhered denatured blood components could be detached by sonication and/or simulated hemodialysis and then gain access into the circulation. The re-entry of the denatured blood components to the patients exposed to reused dialyzers may result in an enhanced immunological response which may contribute to antibody formation (such as anti-Nform antibody) with a reused hemodialyzer.
Insights
Dialyzer reuse may cause adverse reactions. Reprocessed membranes can shed denatured blood components, potentially triggering immune responses like anti-Nform antibody formation in patients undergoing hemodialysis.
Area of Science:
- Biomaterials Science
- Nephrology
- Immunology
Background:
- Dialyzer reuse is common for medical and economic reasons.
- Adverse reactions, including immunohemolytic anemia linked to anti-Nform antibodies, have been reported with reused dialyzers.
- Formaldehyde is a common disinfectant used in dialyzer reprocessing.
Purpose of the Study:
- To investigate the interaction between blood components and reprocessed polysulfone dialyzer membranes.
- To assess the adherence and potential detachment of blood components after formaldehyde disinfection.
- To explore the implications for patient immune responses during hemodialysis.
Main Methods:
- Scanning electron microscopy (SEM) to visualize membrane-blood component interactions.
- Cytologic staining techniques for cellular analysis.
- Sonication and simulated hemodialysis to test detachment of adhered components.
Main Results:
- Various blood components, including fibrin and cells, adhered to the polysulfone membrane post-reprocessing.
- Adhered, denatured blood components could be detached by sonication or simulated hemodialysis.
- Detached components could re-enter the patient's circulation.
Conclusions:
- Reprocessed dialyzers may retain and release denatured blood components.
- The re-entry of these components could enhance immunological responses.
- This may contribute to antibody formation, such as anti-Nform antibodies, in patients using reused hemodialyzers.
Related Concept Videos
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Hemodialysis II: Procedure and Complications
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Hemodialysis I: Introduction
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis