Mechanical stamping: a cause of fatigue fracture
S H Naidu1, C P Warner, C Laird
1Department of Orthopaedic Surgery, Pennsylvania State University, Hershey, PA, USA.
Clinical Orthopaedics and Related Research
|July 1, 1996
Abstract:
Failure analysis of a retrieved flexible intramedullary nail was performed by standard optical and scanning electron microscopy. Results show that the corrosion fatigue at the stamped label caused the nail fracture. The authors strongly recommend that the American Society for Testing of Materials standard F86-91 be modified to exclude mechanical stamping from the repertoire for labeling implants. Surgeons also must avoid using mechanically stamped hardware that will be subject to cyclic loads in vivo.