[Spiral CT with reduced radiation dosage]

C G Diederichs1, H Bruhn, M Funke

  • 1Abteilung Röntgendiagnostik I, Georg-August-Universität Göttingen.

Rofo : Fortschritte Auf Dem Gebiete Der Rontgenstrahlen Und Der Nuklearmedizin
|March 1, 1996
PubMed
Summary

Reduced dose high-resolution spiral-CT (HR-SCT) with multiplanar volume reconstruction (MPVR) significantly lowers radiation exposure for bone imaging. This technique achieves sufficient image quality for routine examinations without diagnostic compromise, except for very small structures.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Publisher Correction: Dimensionally and environmentally ultra-stable polymer composites reinforced with carbon fibres.

Nature materials·2020
Same author

Publisher Correction: Dimensionally and environmentally ultra-stable polymer composites reinforced with carbon fibres.

Nature materials·2020
Same author

Dimensionally and environmentally ultra-stable polymer composites reinforced with carbon fibres.

Nature materials·2019
Same author

Improved survival of Swedish glioblastoma patients treated according to Stupp.

Acta neurologica Scandinavica·2018
Same author

Effectiveness of open and arthroscopic rotator cuff repair (UKUFF): a randomised controlled trial.

The bone & joint journal·2017
Same author

[Diagnostic imaging of breast cancer : An update].

Der Radiologe·2016

Area of Science:

  • Radiology
  • Medical Imaging

Background:

  • High-resolution spiral-CT (HR-SCT) is crucial for detailed imaging of bony structures.
  • Reducing radiation dose in CT scans is a significant clinical goal.

Purpose of the Study:

  • To evaluate the clinical utility of a reduced radiation dose in HR-SCT combined with multiplanar volume reconstruction (MPVR).
  • To assess the effectiveness of this combined technique in decreasing image noise and radiation exposure.

Main Methods:

  • HR-SCT was performed on 33 patients with 1 mm slice thickness, 40 mA, and 80 or 120 kV.
  • 2-D, 3-D, and MPVR image reconstructions were utilized.
  • Skin entrance dose was measured, not exceeding 3 mGy.

Main Results:

Related Experiment Videos

  • Image noise was negligible for high-contrast structures like bones.
  • MPVR effectively reduced image noise in soft tissue imaging.
  • Radiation dose was reduced 10-20 fold, below 3 mGy, without compromising diagnostic quality for most structures.
  • Conclusions:

    • HR-SCT with MPVR allows for significant dose reduction in imaging facial bones, skull, and extremities.
    • The technique is feasible for routine examinations, offering sufficient image quality.
    • Very small bony structures may not be visualized adequately with this reduced dose approach.