Central venous catheter placement using electromagnetic position sensing: a clinical evaluation

H Starkhammar1, M Bengtsson, D A Kay

  • 1Department of Oncology, University of Linköping Hospital, Sweden.

Biomedical Instrumentation & Technology
|March 1, 1996
PubMed
Summary

A new electromagnetic system accurately guided central venous catheter (CVC) placement in 92% of cases, reducing the need for X-ray guidance during the procedure.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Changes in the upper airway volume after orthognathic surgery: three-dimensional measurements in a supine body position.

International journal of oral and maxillofacial surgery·2022
Same author

Complications after minor salivary gland biopsy: a retrospective study of 630 patients from two Swedish centres.

Scandinavian journal of rheumatology·2022
Same author

Mandibulotomy access to tumour sites: fewer complications for postoperative compared with preoperative radiotherapy.

International journal of oral and maxillofacial surgery·2020
Same author

Is there a difference in judgement of facial appearance depending on ethnic background? Photographic evaluation of facial appearance in orthognathic surgery.

The British journal of oral & maxillofacial surgery·2020
Same author

Outcome of photographic evaluation of facial appearance in orthognathic surgery: how does it correlate with planning of treatment and patient-reported outcome?

The British journal of oral & maxillofacial surgery·2019
Same author

A comparison of cost-effectiveness of computer-assisted 2-and 3-dimensional planning techniques in orthognathic surgery.

The British journal of oral & maxillofacial surgery·2019

Area of Science:

  • Medical Devices
  • Interventional Cardiology
  • Radiology

Background:

  • Accurate central venous catheter (CVC) tip placement in the lower superior vena cava is crucial for patient safety.
  • Traditional CVC insertion often relies on anatomical landmarks and fluoroscopy, which carry risks.

Purpose of the Study:

  • To clinically evaluate a novel electromagnetic navigation system for CVC insertion.
  • To assess the accuracy and feasibility of the electromagnetic system in real-world clinical settings.

Main Methods:

  • Fifty central venous catheters were implanted using the electromagnetic system.
  • Fluoroscopy was employed to monitor the accuracy of catheter tip placement by the system.

Main Results:

Related Experiment Videos

  • The electromagnetic system achieved accurate catheter tip placement (within 2.5 cm of the target) in 46 out of 50 cases (92%).
  • Difficulties in determining tip location were noted in two patients with unusual chest anatomy.
  • Two cases had placement errors due to factors unrelated to the electromagnetic system.
  • Conclusions:

    • The electromagnetic system demonstrates high accuracy for CVC placement.
    • This technology offers a potential alternative to blind insertion techniques, reducing reliance on perioperative fluoroscopy.