Related Experiment Video
Updated: Aug 13, 2026

Lung Rapid Recovery Procurement Combined with Abdominal Normothermic Regional Perfusion in Controlled Donation after Circulatory Death
Published on: August 15, 2022
Venous cannula obstruction by Swan-Ganz catheter during cardiopulmonary bypass
J R Oyarzun1, J S Donahoo, J R McCormick
1Section of Cardiothoracic Surgery, University of Medicine and Dentistry-New Jersey Medical School, Newark, USA.
Abstract:
During a routine coronary revascularization procedure in a 51-year-old man, venous return was suddenly interrupted. The venous cannula was rapidly removed, revealing the pulmonary artery catheter wrapped around it. The catheter also had entered the lumen of the cannula, and the inflated balloon completely obstructed the venous return. The prompt removal and reinsertion of the venous cannula resulted in uneventful recovery of the patient.
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Peripheral Artery Disease V: Postoperative Nursing Management