Related Experiment Videos
Thrombogenic properties of untreated and poly(ethylene oxide)-modified polymeric matrices useful for preparing
C Espadas-Torre1, M E Meyerhoff
1Department of Chemistry, University of Michigan, Ann Arbor 48109, USA.
Abstract:
In vitro platelet adhesion studies are used to compare the thrombogenic properties of various polymer matrices useful for preparing implantable ion-selective membrane electrodes. Conventional plasticized poly(vinyl chloride) and alternate polyurethane materials (Tecoflex, Pellethane) doped with proton- (tridodecylamine) and potassium-selective (valinomycin) ionophores are shown to be potentially thrombogenic. Incorporation of high molecular weight block copolymers of poly(ethylene oxide) and poly(propylene oxide) (e.g., Pluronic F108 and Tetronic 1508) within ion-selective membranes reduces platelet adhesion. A more marked decrease in platelet adhesion is, however, observed when the Tecoflex-based membranes are coated with a thin photo-cross-linked layer of poly(ethylene oxide). Such surface-modified membranes are shown to retain potentiometric ion response properties (i.e., selectivity, response times, response slopes, etc.) essentially equivalent to untreated membranes.