Related Experiment Video
Updated: Aug 13, 2026

A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
Direct determination of plasma endothelin-I by chemiluminescence enzyme immunoassay
1Pharmaceutical Research Laboratory, Hitachi Chemical Co., Ibaraki, Japan. HCN00974@niftyserve.or.jp
Abstract:
A highly sensitive sandwich-type chemiluminescence enzyme immunoassay for plasma endothelin-I (ET-1), involving no extraction steps, has been developed. Two populations of polyclonal antibodies were used in the present study: One is specific to the C-terminus of the endothelin family of peptides; the other, a Fab' fragment against the N-terminal core region of ET-1, is coupled with horseradish peroxidase (HRP). Labeled HRP activity was measured by using an enhanced chemiluminescence reaction of luminol/hydrogen peroxide. The assay was sensitive enough to detect 0.5 ng/L (0.05 pg/well) of plasma ET-1 and had no significant cross-reactivities with other related peptides, including endothelin-3 and the endothelin precursor peptide, big ET-1. Reliability of the assay was confirmed by comparison with an extraction-based procedure and a commercial kit.