Related Experiment Videos
New enzymatic assay for calcium in serum
S Kimura1, S Iyama, Y Yamaguchi
1Central Laboratory for Clinical Investigation, Osaka University Hospital, Japan.
Clinical Chemistry
|August 1, 1996
Summary
A new kinetic assay using yeast urea amidolyase offers a simple and rapid method for measuring serum calcium. This assay is reliable, unaffected by common interfering substances, and shows excellent correlation with established methods.
Area of Science:
- Biochemistry
- Clinical Chemistry
- Enzymology
Background:
- Accurate serum calcium measurement is crucial for diagnosing and managing various medical conditions.
- Existing methods may have limitations in terms of speed, complexity, or susceptibility to interference.
- Development of novel, efficient, and reliable calcium assays is an ongoing area of clinical chemistry research.
Purpose of the Study:
- To establish a simple, rapid, and reliable kinetic assay for quantifying calcium in serum.
- To utilize urea amidolyase (EC 3.5.1.45) from yeast species as the core enzyme for calcium detection.
- To validate the performance characteristics and assess the interference profile of the newly developed assay.
Main Methods:
- A kinetic assay was developed based on the inhibitory effect of calcium on yeast urea amidolyase.
- Endogenous ammonium ions were removed using glutamate dehydrogenase (GLDH; EC 1.4.1.4).
- Ammonium ion production, inversely proportional to calcium concentration, was measured spectrophotometrically at 340 nm after NADP+ formation catalyzed by GLDH.
Main Results:
- The assay demonstrated good precision with within-run coefficients of variation (CVs) of 1.7-3.2% and day-to-day CVs of 2.8-4.1%.
- Analytical recovery ranged from 92-112%, indicating accuracy.
- The assay showed excellent correlation with established methods (o-cresolphthalein complexone and another enzymatic method), with correlation coefficients (r) of 0.949 and 0.955, respectively.
Conclusions:
- The developed kinetic assay using urea amidolyase provides a simple, rapid, and accurate method for serum calcium determination.
- The assay is robust, with minimal interference from common serum components and substances.
- This novel enzymatic method offers a reliable alternative for routine clinical laboratory analysis of calcium.