A new miniature x-ray source for interstitial radiosurgery: device description
M Dinsmore1, K J Harte, A P Sliski
1Photoelectron Corporation, Waltham, Massachusetts 02154, USA.
Abstract:
A device that generates low-energy x rays at the tip of a needle-like probe was developed for stereotactic interstitial radiosurgery. Electrons from a small thermionic gun are accelerated to a final energy of up to 40 keV and directed along a 3 mm outside diameter drift tube to a thin Au target, where the beam size is approximately 0.3 mm. All high-voltage electronics are in the probe housing, connected by low-voltage cable to a battery-operated control box. X-ray output, which is nearly isotropic, consists of a bremsstrahlung spectrum and several lines between 7 and 14 keV, with characteristic radiation contributing 15% of the total energy output. To date, 14 patients with metastatic brain tumors have been treated with this device.