Effects of electrode polarity on defibrillation thresholds in biphasic endocardial defibrillation
J Neuzner1, H F Pitschner, T Schwarz
1Department of Cardiology, Kerckhoff-Clinic, Bad Nauheim, Germany.
The American Journal of Cardiology
|July 1, 1996
Abstract:
The defibrillation thresholds of both first-phase polarities of a biphasic waveform were tested using an endocardial defibrillation electrode system. Despite differences in defibrillation thresholds in the individual patient, both tested electrode polarities lead to a comparable overall defibrillation efficacy.