Defibrillation shocks over epicardial patches produce sympathetic neural dysfunction in man
L B Rigden1, R D Mitrani, H N Wellman
1Indiana University School of Medicine, Department of Medicine, Krannert Institute of Cardiology Indianapolis 46202-4800, USA.
Introduction:
The purpose of this study was to determine the effect of direct current (DC) shocks on cardiac sympathetic innervation in humans using I-123-metaiodobenzylguanidine (MIBG) scintigraphy. Decreased efferent sympathetic neural function has been demonstrated following > 10-J DC shocks delivered through epicardial patch electrodes in dogs. To evaluate the effect of DC shocks on cardiac sympathetic innervation in humans, we performed MIBG scintigraphy in 11 patients (ages 46 to 75 years) prior to and after receiving shocks from an implantable cardioverter defibrillator (ICD).
Methods And Results:
This study was performed during an ICD generator change in 7 patients with epicardial patch electrodes and at the time of initial ICD implantation in 4 patients: 2 with epicardial patch electrodes, and 2 with a transvenous ICD system. All patients had spontaneous and inducible ventricular tachycardia. Prior to ICD implantation and remote from any cardioversions or shocks, baseline MIBG and thallium-201 scintigraphy were performed. Repeat MIBG scintigraphy was performed after delivery of ICD shocks and compared with the baseline scans to determine the effect of the shock on sympathetic neural function. The baseline scans revealed focal areas of reduced MIBG uptake in areas of thallium perfusion defects in all patients except the patient without structural heart disease whose scans were normal. Postshock, patients with epicardial patch electrodes who received at least one 24-J shock and had the postshock MIBG scan performed within 4 hours demonstrated no cardiac uptake of MIBG. Two patients with epicardial patch electrodes had no change in the postshock MIBG scans: 1 had a maximal shock of 20 J, and the other had the postshock scan delayed for 11 hours. The 2 patients with a transvenous lead system demonstrated no change in the postshock MIBG scan when compared with baseline.
Conclusions:
This study demonstrates that following DC shocks delivered over epicardial patch electrodes, there is diffuse reduction in MIBG uptake that probably represents cardiac sympathetic neural dysfunction that appears to be transient. Sympathetic function does not appear to be affected by shocks delivered over a transvenous lead system.
More Related Videos
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
08:56Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias