Control of intrauterine device insertion with three-dimensional ultrasound: is it the future?
F Bonilla-Musoles1, F Raga, N G Osborne
1Department of Obstetrics and Gynecology, University of Valencia School of Medicine, Spain.
Abstract:
A comparative study was designed to determine whether three-dimensional transvaginal sonography (3D-TVS) offered advantages over two-dimensional transvaginal sonography (2D-TVS) for the identification and location of IUDs in 66 asymptomatic women. Hysteroscopy was performed in cases in which there was a discrepancy between the information obtained by both methods (n = 14). In eight cases (12.2%) the IUD was misidentified with 2D-TVS. In six cases (9.1%) it was not possible to identify the device model with 2D-TVS. In two cases (3.0%) 2D-TVS failed to identify the position of the device. In contrast, all IUDs were identified and located accurately with 3D-TVS.