Background:
Electrophrenic pacing can be used in the management of ventilatory failure in quadriplegic patients. A study was undertaken to determine the pattern of transdiaphragmatic pressure (PDI) during the conditioning phase of electrophrenic pacing to see if it had a possible role in optimising the process of conditioning.
Methods:
The tidal volume (TV) and PDI were measured in a group of six quadriplegic patients commencing ventilation by low frequency pulse stimulation (7-10 Hz) and low respiratory rate stimulation (< 10 breaths/min).
Results:
Tidal volume increased between baseline and month 1 (4.33 ml/kg, p < 0.001) and between months 1 and 2 (3.00 ml/kg, p < 0.05) and then stabilised. PDI was higher during bilateral diaphragmatic pacing (mean (SD) 1.73 (0.30) kPa) than with either left (1.15 (0.34) kPa) or right (0.86 (0.37) kPa) unilateral pacing. PDI varied throughout the observation period, probably by interaction between recovery of the diaphragmatic fibres and the pacing regimen.
Conclusions:
Patients with quadriplegia due to high spinal injury can be maintained with ventilation by continuous electrophrenic pacing. The control criteria used in this study for pacing were tidal volume and the patient's tolerance, and the PDI measurement did not contribute any additional information to help with managing the conditioning process.
More Related Videos
Related Concept Videos
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...