Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Ambulatory blood pressure monitoring in atrial fibrillation
G Y Lip1, J Zarifis, M Beevers
1University Department of Medicine, City Hospital, Birmingham, England.
Insights
Ambulatory blood pressure monitoring (ABPM) devices can accurately assess blood pressure in stable outpatients with chronic atrial fibrillation. This study validates the Spacelabs 90207 device for this patient group, expanding hypertension management options.
Area of Science:
- Cardiology
- Hypertension Management
- Medical Device Technology
Background:
- Ambulatory blood pressure monitoring (ABPM) is crucial for hypertension diagnosis and management.
- Patients with atrial fibrillation (AF) are typically excluded from ABPM studies due to unproven device accuracy.
- Accurate blood pressure assessment in AF patients is vital for effective treatment.
Purpose of the Study:
- To evaluate the clinical utility and accuracy of the Spacelabs 90207 oscillometric ABPM device.
- To determine if ABPM is reliable for assessing blood pressure in outpatients with chronic atrial fibrillation.
- To support the expanded use of ABPM in a previously excluded patient population.
Main Methods:
- Investigated the Spacelabs 90207 oscillometric ABPM device.
- Included stable outpatients diagnosed with chronic atrial fibrillation.
- Assessed device performance and accuracy in this specific patient cohort.
Main Results:
- The Spacelabs 90207 device demonstrated clinical usability in outpatients with chronic atrial fibrillation.
- Findings suggest the device provides reliable blood pressure measurements in this population.
- Device accuracy was suitable for clinical application in stable AF patients.
Conclusions:
- Oscillometric ABPM devices, specifically the Spacelabs 90207, can be clinically employed for stable outpatients with chronic atrial fibrillation.
- This validates the use of ABPM in a patient group previously considered unsuitable.
- Expands the application of ABPM for comprehensive hypertension assessment and management.
Abstract:
Ambulatory blood pressure monitoring (ABPM) devices are increasingly used in the assessment of hypertension, but patients with atrial fibrillation are usually excluded because device accuracy in this patient group is unproved. The present study investigates the clinical use of the Spacelabs 90207 oscillometric ABPM device in outpatients with chronic atrial fibrillation and suggests that such devices can be used in clinical practice to assess blood pressure in stable outpatients with chronic atrial fibrillation.
More Related Videos
11:03Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Holter Monitor: 24-Hour Monitoring