Chamber for indirect calorimetry with improved transient response

B Henning1, R Löfgren, L Sjöström

  • 1Department of Medicine, Sahlgrenska Hospital, Göteborg, Sweden.

Medical & Biological Engineering & Computing
|May 1, 1996
PubMed
Summary

A new indirect calorimetry chamber was developed with advanced algorithms for precise respiration calculations. This system effectively suppresses noise and identifies trends, enabling rapid responses for energy expenditure studies.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Vemurafenib and cobimetinib combination therapy for BRAFV600E-mutated melanoma favors posterior reversible encephalopathy syndrome.

Annals of oncology : official journal of the European Society for Medical Oncology·2019
Same author

Body composition through adult life: Swedish reference data on body composition.

European journal of clinical nutrition·2014
Same author

The incidence of albuminuria after bariatric surgery and usual care in Swedish Obese Subjects (SOS): a prospective controlled intervention trial.

International journal of obesity (2005)·2014
Same author

Search for dinucleon decay into kaons in Super-Kamiokande.

Physical review letters·2014
Same author

Hormones, body composition and cardiovascular risk.

Asia Pacific journal of clinical nutrition·2014
Same author

Carotid artery remodelling in relation to body fat distribution, inflammation and sustained weight loss in obesity.

Journal of internal medicine·2013

Area of Science:

  • Physiology
  • Biomedical Engineering
  • Respiratory Medicine

Background:

  • Indirect calorimetry chambers require accurate respiration calculations.
  • Large chambers present challenges due to small gas concentration changes.
  • Existing methods may lack sensitivity for rapid physiological changes.

Purpose of the Study:

  • To construct and validate a novel indirect calorimetry chamber.
  • To develop algorithms for enhanced noise suppression and trend identification in gas analysis.
  • To enable precise respiration measurements even with minimal gas concentration variations.

Main Methods:

  • Construction of a large-volume indirect calorimetry chamber.
  • Development of algorithms based on the exact steady-state solution of respiration equations.

Related Experiment Videos

  • Application of a least-squares method fitting gas concentrations to two exponential segments.
  • Evaluation of gas concentration and its derivative at -15 min, repeated every minute.
  • Main Results:

    • The developed algorithms provide effective noise suppression and trend identification.
    • Gas injection tests confirmed an instantaneous response to single respiration changes.
    • The system correctly averages repeated respiration changes within 15-minute periods.
    • The procedure demonstrates high accuracy and rapid responsiveness.

    Conclusions:

    • The new indirect calorimetry chamber is suitable for traditional 24-hour energy expenditure measurements.
    • The chamber's rapid response capability makes it valuable for experiments requiring dynamic physiological monitoring.
    • The advanced algorithms enhance the precision and utility of indirect calorimetry.