[Problems and limits in noninvasive oxygen monitoring]
Related Experiment Video
Updated: Oct 7, 2026

Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
1Klinik für Anästhesiologie der Universität Heidelberg.
Abstract:
Recently, professional anaesthesiologists organizations, have set up recommendations for intraoperative monitoring. These recommendations are based upon the assumption that anaesthesia-related deaths are largely preventable and that extensive monitoring will reduce mortality. Hypoxaemia appears to be an important pathophysiological mechanism, and this is why oxygen monitoring is given a high priority in safety standards in anaesthesia. However, no reduction in mortality has been definitively shown to follow the use of any monitoring device in randomized, controlled trials. Four techniques of oxygen monitoring are discussed and evaluated. Oxygen monitoring during inspiration can be recommended, although its implementation will probably prevent only a few anaesthesia-related deaths. Pulse oximetry is certainly highly recommended, but also has some deficiencies which are not sufficiently well known considering the popularity of this technique. Transcutaneous PO2 is considered mainly for historical reasons. Near infrared spectrometry is a new technique offering noninvasive and continuous monitoring of intracerebral oxygenation.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...