Comparison of Chemcard cholesterol test and laboratory cholesterol measurements

I E Gan1, G M Schier

  • 1Biochemistry Department, Illawarra Regional Hospital, Wollongong, NSW.

Australian and New Zealand Journal of Medicine
|December 1, 1995
PubMed

Insights

The Chemcard cholesterol test shows poor agreement with laboratory results, potentially misclassifying patient risk. Formal laboratory cholesterol measurement remains necessary.

Area of Science:

  • Clinical Chemistry
  • Diagnostic Accuracy
  • Cardiovascular Risk Assessment

Background:

  • The Chemcard cholesterol test is intended for preliminary cholesterol level estimation in general practice settings.
  • Elevated levels detected by Chemcard may prompt referral for laboratory confirmation.
  • This study evaluates the diagnostic performance of the Chemcard test.

Purpose of the Study:

  • To compare the accuracy of the Chemcard cholesterol test against a standardized laboratory procedure.
  • To assess the reliability of Chemcard for estimating cholesterol levels.

Main Methods:

  • The Chemcard cholesterol procedure was applied to 200 community subjects.
  • Participants were enrolled in a cardiovascular risk assessment program.
  • Paired measurements from Chemcard and laboratory methods were analyzed.

Main Results:

  • A correlation coefficient of 0.83 was observed between Chemcard and laboratory cholesterol measurements.
  • Chemcard results were slightly higher than laboratory results (mean difference: 0.16 +/- 0.58 mmol/L).
  • Significant risk group misclassification was identified.

Conclusions:

  • The Chemcard cholesterol test demonstrates poor agreement with laboratory methods.
  • The test's potential for misclassification suggests it is not a reliable standalone system.
  • Formal laboratory cholesterol measurement is still required, refuting manufacturer claims.