Related Experiment Video
Updated: Jul 20, 2026

Continuous-wave Thulium Laser for Heating Cultured Cells to Investigate Cellular Thermal Effects
Published on: June 30, 2017
Erbium laser ablation of dental hard tissue: effect of water cooling
S R Visuri1, J T Walsh, H A Wigdor
1Biomedical Engineering Department, Northwestern University, Evanston, Illinois 60208, USA.
Background And Objective:
Several lasers have been explored for hard dental tissue applications; used alone they have resulted in potentially harmful temperature increases in the pulp chamber.
Materials And Methods:
An Er:YAG laser (lambda = 2.94 microns) was used to ablate hard dental tissues. Ablation rates with and without a water-cooling spray were measured. Subsequent experiments investigated the cooling effects of the water. Initially single channels were drilled into dentin; further studies involved ablating rectangular areas with repetition rates up to 10 Hz.
Results:
The water spray minimally reduced the ablation rates of dentin and did not affect the ablation rates of enamel. The water spray effectively cooled the teeth; while using the maximum average power investigated (10 Hz, 360 mJ/pulse), a water flow rate of 4.5 ml/min limited the temperature rise in the pulp chamber to less than 3 degrees C.
Conclusion:
The studies confirm the feasibility of using an Er:YAG laser in conjunction with a water spray to safely and effectively remove hard dental tissues.