Related Experiment Videos
Pulsatile perfusion system for ex vivo investigation of biochemical pathways in intact vascular tissue
R F Labadie1, J F Antaki, J L Williams
1Department of Surgery, University of Pittsburgh, Pennsylvania 15261, USA.
Abstract:
We have constructed and performed initial validation of an innovative perfusion system that allows exposure of intact segments of vascular tissue to realistic physiological and hemodynamic environments ex vivo. Computer-controlled opening and closing of an in-line gate valve allows generation of arterial pressure waveforms. The control algorithm predicted resultant pressure waveforms with a high degree of accuracy (Pearson correlation coefficient > 0.97). To document vascular homeostasis ex vivo, vasomotor bioassays and morphological studies were performed. The bioassays consisted of injecting epinephrine (2 x 10(-3) mg/ml) into the perfusion system followed by acetylcholine (100 microM) while concurrently measuring vessel diameter with a laser micrometer, significant vasomotion was measured for canine carotid arteries (n = 4) bioassayed after 1, 24, and 48 h of perfusion (P < 0.03). Additionally, human saphenous vein segments were perfused for 24 h (n = 4) and viewed with laser confocal scanning microscopy and transmission electron microscopy; photomicrographs show typical vascular morphology. We conclude that the vascular perfusion system described herein is well suited for investigating the response of intact vascular tissue to hemodynamic variables.