Related Experiment Video
Updated: Jul 31, 2026

One-channel Cell-attached Patch-clamp Recording
Published on: June 9, 2014
Dependence of solution exchange time on cell or patch linear dimensions in concentration jump experiments using
1Department of Neurophysiology, Federal Institute for Neurobiology, PO Box 1860, Brenneckestrasse 6, D-39008 Magdeburg, Germany.
Abstract:
A critical evaluation of "concentration jump" techniques employing hydrodynamic solution exchange has been undertaken. Data obtained from experiments using acutely isolated sensory neurones demonstrated a dependence of the total solution exchange time on the linear dimensions of the preparation. The exchange time around a circular surface is proportional to the radius of this surface. A quantitative approximation can be formulated as follows: if the membrane area of the object under investigation is increased n times, the duration of the full solution exchange is increased by radical n times. A nonlinear dependence of exchange time on vacuum pressure was confirmed in a separate set of experiments.