Flow continuity of infusion systems at low flow rates
1Perinatal Research Centre, Royal Women's Hospital, Brisbane, Queensland.
Anaesthesia and Intensive Care
|October 1, 1995
Summary
Neonatal intensive care requires precise low-rate infusions. Some syringes provide continuous flow for inotropes, while certain infusion pumps do not, ensuring safer neonatal hemodynamic management.
Area of Science:
- Neonatal intensive care
- Medical device engineering
- Pharmacology
Background:
- Infusion of fluids and drugs at very low rates is critical in neonatal intensive care.
- Hemodynamic fluctuations during inotrope infusion are linked to syringe driver plunger sticking.
- Regulatory bodies recommend specific pumps for these critical infusions.
Purpose of the Study:
- To evaluate the performance of various infusion systems for low-rate drug delivery in neonatal care.
- To identify infusion devices suitable for continuous and safe administration of inotropes.
- To assess the reliability of syringe drivers and infusion pumps at low flow rates.
Main Methods:
- Testing of multiple syringe driver and infusion pump systems.
- Assessment of flow rate continuity at very low infusion rates.
- Evaluation of device suitability for inotrope delivery based on flow characteristics.
Main Results:
- Certain syringes demonstrated continuous flow at low rates, making them suitable for inotropes.
- Some tested infusion pumps exhibited non-continuous flow at low rates, posing a risk.
- Device performance varied significantly, impacting the safety of neonatal infusions.
Conclusions:
- The choice of infusion device is critical for maintaining stable hemodynamics in neonates.
- Syringes providing continuous flow are preferable for low-rate inotrope infusions.
- Infusion pumps with non-continuous low-rate flow are unsatisfactory for this application.