Related Experiment Videos
CAP interlaboratory survey of analytic balances
American Journal of Clinical Pathology
|July 1, 1977
Summary
An interlaboratory survey found that analytic balances accurately weigh samples. For 525 mg samples, weighing errors rarely exceed +/-0.4%, ensuring reliable laboratory measurements.
Area of Science:
- Analytical Chemistry
- Metrology
Background:
- Accurate weighing is fundamental in analytical chemistry and metrology.
- Variability in analytic balance performance can impact experimental results.
Purpose of the Study:
- To assess the performance and accuracy of analytic balances across multiple laboratories.
- To quantify the typical error margins for weighing small samples.
Main Methods:
- An interlaboratory survey was conducted.
- Aluminum oxide spheres (approx. 525 mg) were used as test samples.
- Data from 83 participating laboratories were analyzed.
Main Results:
- 82 of 83 laboratories completed the survey.
- Results indicate that for 525 mg samples, errors typically do not exceed +/-0.3% in 95% of weighings (19/20).
- Errors are expected not to exceed +/-0.4% in 99.7% of weighings (997/1000).
Conclusions:
- Analytic balances demonstrate high accuracy and reliability at the 525 mg weight level.
- The survey provides statistical data on expected weighing error margins.
- Understanding variance components is crucial for maintaining measurement precision.