Related Experiment Video
Updated: Aug 15, 2026

Dried Blood Spots - Preparing and Processing for Use in Immunoassays and in Molecular Techniques
Published on: March 13, 2015
A rapid anion exchange high-performance liquid chromatography method for the measurement of HbA2 in whole blood
B G Keevil1, P W Maylor, D Rowlands
1Department of Clinical Chemistry, South Manchester University Hospitals NHS Trust, Wythenshawe Hospital.
Abstract:
We developed a binary gradient high-performance liquid chromatography (HPLC) method for measuring HbA2 in whole blood samples using a Pharmacia Mono Q column (1 mL) and measurement at 415 nm. The assay requires a simple lysis and centrifugation step before injection onto the column. We found good agreement of results between the HPLC method and the Helena column chromatography method. The within batch precision was 2-6% and between batch precision was 4-6%. We found that using 30 mM Tris buffers (pH 7-8) with a sodium chloride gradient resulted in short analysis times and good chromatographic separation of HbA2, HbS and HbA. We conclude that this is a robust assay for the diagnosis of beta-thalassaemia.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors
Affinity Chromatography