Statistics for critical clinical decision making based on readings of pairs of implanted sensors
D W Schmidtke1, M V Pishko, C P Quinn
1Department of Chemical Engineering, University of Texas at Austin 78712-1062, USA.
Abstract:
Low error rates are essential if lives of patients are to depend on readings of implanted sensors, such as glucose sensors in insulin-dependent diabetic patients. To verify the operation and to calibrate on demand an implanted sensor, it is necessary that calibration through a single, independent measurement involving withdrawal of only one sample of blood and its independent analysis be feasible. Such a one-point calibration must be accurate. Borrowing from nuclear reactor safety assurance, where a likelihood ratio test is applied to readings of pairs of pressure sensors for shutdown/no shutdown decisions, we apply a similar test to sensor pairs implanted in rats. We show, for five sets of glucose sensor pairs, calibrated in vivo by withdrawal of a single sample of blood, that application of the likelihood ratio test increases the fraction of the clinically correct readings from 92.4% for their averaged readings to 98.8%.