Related Experiment Video
Updated: Feb 8, 2026

Quantification of Optic Nerve Cross Sectional Area on MRI: A Novel Protocol using Fiji Software
Published on: September 4, 2021
Evaluation of optic disc measurements with the glaucoma-scope
J A Dan1, D A Belyea, M F Lieberman
1Ophthalmology Service, Walter Reed Army Medical Center, Washington, DC 20307-5001, USA.
The Glaucoma-Scope shows moderate reproducibility for optic disc analysis in select cases. However, its diagnostic utility for glaucoma is questionable due to poor agreement with expert evaluation and image acquisition challenges.
Area of Science:
- Ophthalmology
- Medical Imaging
- Glaucoma Diagnostics
Background:
- Glaucoma diagnosis relies on accurate optic disc evaluation.
- New imaging technologies aim to improve diagnostic capabilities.
- The Glaucoma-Scope is a device for optic disc imaging and analysis.
Purpose of the Study:
- To assess the Glaucoma-Scope's image acquisition success rate.
- To determine the reproducibility of Glaucoma-Scope analysis.
- To compare Glaucoma-Scope findings with expert evaluations.
Main Methods:
- 168 patients underwent Glaucoma-Scope analysis.
- Image acquisition failures and associated conditions were recorded.
- Inter- and intraobserver reproducibility was calculated using weighted Kappa and coefficient of variation.
- Glaucoma-Scope cup-to-disc ratio was compared to expert assessment.
Main Results:
- 14% of eyes (336 total) were unsuitable for Glaucoma-Scope analysis due to various ocular conditions.
- Quantitative analysis showed low variation and moderate-to-substantial inter- and intraobserver agreement.
- Significant disagreement was found between Glaucoma-Scope and expert cup-to-disc ratio evaluations (weighted K = -0.11).
Conclusions:
- The Glaucoma-Scope demonstrates moderate to substantial reproducibility in optic disc topography analysis under specific conditions.
- Ocular conditions like cataracts, aphakia, and pigmented fundi limit its analyzability.
- Low agreement with expert evaluation questions its current diagnostic value for glaucoma, though it may aid in change detection with optimal imaging.
Related Concept Videos
Design Example: Calculating Safe Diameter for Wind-Exposed Disc
The Scope of Physics
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Self-Evaluation: Self-Enhancement and Self-Verification