Clinical evaluation of a multiparameter intra-arterial blood-gas sensor
E Abraham1, T J Gallagher, S Fink
1Division of Pulmonary Sciences and Critical Care Medicine, University of Colorado Health Sciences Center, Denver 80262, USA.
Objective:
To evaluate the in vivo performance of a continuous, intra-arterial, multiparameter blood-gas sensor containing a thermocouple, miniaturized polarographic oxygen electrode, as well as fiberoptic pH and pCO2 sensors.
Design:
Prospective, multicenter study comparing pH, PaCO2, and PaO2 measurements from the intraarterial sensor with those obtained from a conventional, laboratory blood-gas monitor.
Setting:
Intensive care units in three academic medical centers.
Patients:
Adult ICU patients (n = 26) with a clinical need for a radial artery catheter and frequent monitoring of arterial blood gases for > or = 3 days.
Interventions:
All patients had the multiparameter intra-arterial sensor placed through a 20-gauge catheter into the radial artery. Every 4 h, or more frequently if clinically indicated, a blood-gas sample was withdrawn from the radial artery catheter and sent to the laboratory for analysis using a conventional laboratory blood-gas analyzer. Immediately prior to withdrawal of the arterial blood, values for pH, PaCO2, and PaO2 from the multiparameter intra-arterial sensor were recorded.
Measurements And Main Results:
The multiparameter sensor was placed into 26 patients. In 7 of the patients, premature discontinuation of monitoring was necessary because of dampening of the pressure tracing, difficulty in withdrawing blood from the arterial catheter, or bending of the extra-arterial fiberoptic channel owing to the cable being inadequately secured (mean monitoring time in these 7 patients: 40.6 +/- 25.7 h). In the other 19 patients, monitoring was continued until no longer clinically indicated. In these patients, a total of 341 data sets was collected, with the average length of monitoring being 69.9 +/- 37.9 h (range 22.0 to 57.9 h). Comparison of the sensor values with those from the blood-gas analyzer showed bias and precision values of 0.006 and 0.026 for arterial pH - 1.19% and 12.54% for PaO2, and 1.28 mmHg and 2.48 mmHg for PaCO2, respectively. No complications were associated with the intra-arterial sensor.
Conclusions:
Clinical performance of this intra-arterial, multiparameter blood-gas sensor demonstrated stability, consistency, and accuracy comparable to laboratory blood-gas analyzers. The present multiparameter, intravascular blood-gas sensor, when inserted in the radial artery, can provide stable and accurate monitoring of pH, PaCO2, and PaO2 over clinically relevant periods of as long as 6 days in the critical care setting.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and HCO3− values are examined to...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation