Eccentric and concentric isokinetic knee flexion and extension: a reliability study using the Cybex 6000 dynamometer
R C Li1, Y Wu, N Maffulli
1Hong Kong Centre of Sports Medicine and Sports Science, Hong Kong.
British Journal of Sports Medicine
|June 1, 1996
Summary
The Cybex 6000 isokinetic dynamometer demonstrates high reliability for assessing knee muscle performance, both concentrically and eccentrically. Peak torque measurements showed the highest reliability, with values generally improving at faster testing speeds.
Area of Science:
- Biomechanics
- Sports Medicine
- Rehabilitation Science
Background:
- Isokinetic dynamometry is crucial for evaluating muscle function.
- Assessing the reliability of specific devices like the Cybex 6000 is essential for accurate clinical and research applications.
- Understanding knee muscle performance is vital for injury prevention and rehabilitation.
Purpose of the Study:
- To evaluate the test-retest reliability of the Cybex 6000 isokinetic dynamometer.
- To assess the reliability of concentric and eccentric knee muscle performance measurements.
- To compare the reliability of different performance variables (peak torque, total work, average power) and testing speeds.
Main Methods:
- Thirty healthy subjects (18 male, 12 female) underwent isokinetic testing of knee flexor and extensor muscles.
- Testing was performed at 60°/s and 120°/s using a continuous concentric-eccentric protocol.
- Interclass correlation coefficient (ICC 2,1) was used to determine reliability (P < 0.05).
Main Results:
- High test-retest reliability was found for peak torque (ICC 0.82–0.91), total work (ICC 0.76–0.89), and average power (ICC 0.71–0.88).
- Reliability was significantly higher at 120°/s compared to 60°/s for all variables.
- Knee extensors and concentric contractions demonstrated greater reliability.
Conclusions:
- The Cybex 6000 is a reliable tool for measuring concentric and eccentric knee muscle performance.
- Clinicians and researchers should consider potential minor fluctuations in muscle performance measurements between tests.
- Higher testing speeds (120°/s) and concentric modes yield more reproducible results.