Method development for the determination of cefdinir and its related substances by high-performance liquid
Abstract:
The validation of the HPLC method used for the determination of cefdinir and its related substances is described. The developed method was specific and stability-indicating and provided a linear response with concentration. The system and method precision, expressed as relative standard deviations, were not greater than 1%, and the reproducibilities within and between laboratories were acceptable for the assay method. The procedure can quantitate related substances greater than approximately 0.05% of the principal cefdinir peak.
More Related Videos
11:17Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
08:54Determining Four Components in a Lipid Nanoparticle RNA Delivery System by Liquid Chromatography Combined with Evaporative Light Scattering Detector
Published on: May 30, 2025
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors