Automated quantitative determination of a new polyamine biosynthesis inhibitor (CGP 48 664) and a potential metabolite in human and animal plasma by high-performance liquid chromatography
Related Experiment Videos
Abstract:
A specific and sensitive liquid chromatographic assay for the determination of 4-amidino-1-indanone-2'-amidinohydrazone (CGP 48 664, I) and a potential metabolite, 2-(4-carbamoyl-2,3-dihydro-1H-inden-1-yliden) hydrazine carboximidamide (CGP 53 391, II), in human and animal plasma was developed. CGP 51 467, a structural analog, was added to the plasma samples (up to 1 ml) as an internal standard. After mixing, the samples were processed automatically using an ASPEC solid-phase extraction system. The final extracts were chromatographed on a 5 microns Purospher RP-18 HPLC column. Chromatography was performed using a gradient system and UV detection. The described HPLC method is suitable for specific and quantitative measurement of concentrations of I, as well as its potential metabolite II down to 5-10 ng/ml in human and animal (dog, rat) plasma with acceptable reproducibility and accuracy.