Verification of absorbed dose determined with plane-parallel chambers in clinical electron beams following AAPM Task
Z Xu1, H Li, P R Almond
1Department of Radiation Oncology, University of Louisville, Brown Cancer Center, Kentucky 40202, USA.
Abstract:
The absorbed dose values determined with the Exradin and PTW-Markus plane-parallel chambers were compared to the values obtained with the ferrous sulphate dosimetry for a number of the Philips SL25 and the Therac 20 electron beams. For the plane-parallel chambers, the cavity-gas calibration factor Ngaspp, was derived by a direct comparison with a calibrated cylindrical chamber using the three different calibration methods as proposed by the newly published AAPM TG 39 protocol. For the ferrous sulphate dosimetry, an epsilon mG value of 352 x 10(-6) m-2 kg-1 Gy-1 was adopted from ICRU Report No. 35. The average ratio of the dose values determined with the plane-parallel chambers and the dose values determined with the Fricke dosimetry system was 1.001 +/- 1.4%. These measurements are consistent with the AAPM TG 39 protocol.