Study of secondary flow in centrifugal blood pumps using a flow visualization method with a high-speed video camera
Related Experiment Video
Updated: Nov 4, 2026

Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
1Department of Applied Electronic Engineering, Faculty of Science and Engineering, Tokyo Denki University Saitama, Japan.
Abstract:
Four pump models with different vane configurations were evaluated with flow visualization techniques using a high-speed video camera. These models also were evaluated through in vivo hemolysis tests using bovine blood. The impeller having the greatest fluid velocity relative to the impeller, the largest velocity variance, and the most irregular local flow patterns in the flow passage caused the most hemolysis. Even if the pumps were operated at almost the same speed (rpm) at the same output, the impeller showing more irregular flow patterns had a statistically greater rate of hemolysis. This fact confirms that the existence of local irregular flow patterns in a centrifugal blood pump deteriorates its hemolytic performance. Thus, to optimize the design of the pump, it is very important to examine the secondary flow patterns in the centrifugal blood pump in detail using flow visualization with a high-speed video camera.
Related Concept Videos
Rapidly Varying Flow
Steady, Laminar Flow in Circular Tubes
Laminar and Turbulent Flow
Eulerian and Lagrangian Flow Descriptions
The Eulerian method focuses on fixed points in space where fluid properties, such as velocity, pressure, and temperature, are observed as the fluid moves between these...
Bernoulli's Equation for Flow Normal to a Streamline
The pressure difference depends on the fluid's velocity and radius of curvature. The pressure variation is minimal in flows with nearly straight streamlines. However, the...