Oversensing of atrial flutter in the recipient atrium of a heart transplant patient with a permanent atrial pacemaker
D A Woodard1, J B Conti, A B Curtis
1Department of Medicine, University of Florida, Gainesville 32610, USA.
Clinical Cardiology
|July 1, 1996
Abstract:
Sinus node dysfunction requiring permanent pacemaker implantation is a common problem following heart transplantation. There are unique problems to consider in pacing the transplanted heart. This report describes a case of far-field sensing of atrial flutter in the recipient atrial remnant, with resultant inhibition of the atrial pacemaker in the donor atrium, and symptomatic bradycardia. Although postulated to be a potential problem, oversensing of the recipient atrium has not been reported prior to this case. Awareness of this problem will provide for more effective pacemaker troubleshooting.