Isolated mitral replacement with stent-mounted antibiotic-treated aortic allograft valves
The Journal of Thoracic and Cardiovascular Surgery
|August 1, 1977
Summary
This study evaluated stent-mounted antibiotic-treated aortic allograft valves for mitral valve disease. High complication rates, including valve failure and reoperation, led to the discontinuation of this device in mitral valve replacement.
Area of Science:
- Cardiac Surgery
- Biomaterials Science
- Clinical Outcomes Research
Background:
- Mitral valve disease affects a significant patient population, often requiring surgical intervention.
- Previous valve replacement strategies have limitations, necessitating novel approaches.
- Stent-mounted allograft valves offer a potential alternative for mitral valve repair.
Purpose of the Study:
- To report the outcomes of mitral valve replacement using stent-mounted antibiotic-treated aortic allograft valves.
- To assess the safety and efficacy of this specific valve prosthesis in a clinical setting.
- To identify complications and failure modes associated with this device.
Main Methods:
- A cohort of 129 patients with isolated mitral valve disease underwent valve replacement.
- The prosthesis used was a stent-mounted antibiotic-treated aortic allograft valve.
- Patients were followed to assess hospital mortality, late mortality, reoperation rates, and valve-related complications.
Main Results:
- Hospital mortality was 3.9%.
- The cumulative complication-free rate at 5 years was only 37%.
- Major complications included late death (21%), reoperation (15%), embolic episodes (4%), and significant valve incompetence (4%).
- Valve failure was primarily due to detachment of the aortic wall remnant from the stent, with a 16% incidence at 5 years.
Conclusions:
- The stent-mounted antibiotic-treated aortic allograft valve demonstrated a high rate of complications in the mitral position.
- Detachment of the valve from the stent was a significant failure mechanism.