Plasma catecholamine assays: calibration with spiked plasma versus aqueous solutions
M Candito1, F Bree, A M Krstulovic
1Laboratoire de Biochimie, Hôpital Pasteur, Nice, France.
Biomedical Chromatography : BMC
|January 1, 1996
Summary
Spiked plasma samples may yield inaccurate results for catecholamine assays due to protein binding. Using aqueous standards for calibration is preferred for reliable free catecholamine measurements in plasma.
Area of Science:
- Clinical Chemistry
- Analytical Biochemistry
- Pharmacology
Background:
- Plasma constituent assays commonly use both aqueous standards and spiked plasma.
- Previous high-performance liquid chromatographic (HPLC) assays using spiked plasma for norepinephrine, epinephrine, and dopamine showed lower detection thresholds, indicating higher concentrations.
- This suggests potential interference or binding of spiked catecholamines during the assay process.
Purpose of the Study:
- To investigate the hypothesis that spiked catecholamines bind to plasma proteins, affecting extraction efficiency prior to HPLC.
- To compare the recovery of catecholamine standards in human albumin solution versus aqueous standards.
- To determine the optimal calibration method for accurate free catecholamine assays in plasma.
Main Methods:
- Comparison of catecholamine recovery rates using standards dissolved in human albumin solution.
- Comparative analysis with recovery rates using aqueous standards.
- High-performance liquid chromatography (HPLC) was employed for the analysis of catecholamines.
Main Results:
- Albumin was identified as a significant factor contributing to the reduced recovery of catecholamines.
- The binding of catecholamines to albumin interferes with their extraction process.
- Extraction of aqueous standards demonstrated a more reliable calibration method compared to using spiked plasma.
Conclusions:
- Albumin in plasma significantly impacts the recovery of spiked catecholamines, leading to inaccurate assay results.
- Calibration using aqueous standards is superior to using spiked plasma for accurate free catecholamine measurements.
- This study highlights the importance of considering protein binding in the development and validation of bioanalytical methods for catecholamines.