Perception of internally generated noise in hearing amplification
1Starkey Laboratories, Eden Prairie, Minnesota, USA.
Journal of the American Academy of Audiology
|August 1, 1996
Summary
Hearing aid amplifier noise can be distracting. This study identified the levels at which internal noise becomes audible and objectionable for individuals with hearing loss, aiding hearing aid fitting.
Area of Science:
- Audiology
- Acoustics
- Hearing Science
Background:
- Internal noise in hearing aid amplifiers can impair speech understanding.
- Determining audible and objectionable noise levels is crucial for hearing aid specification and fitting.
Purpose of the Study:
- To determine the perceived audible and objectionable levels of internal hearing aid amplifier noise.
- To compare noise perception between individuals with hearing loss and normal hearing.
- To investigate the relationship between noise levels, audiometric configuration, and noise perception.
Main Methods:
- Eight subjects with moderate hearing loss and five normal-hearing subjects were tested.
- A test amplifier without acoustic input was used to generate internal noise.
- Two noise types were simulated: output stage noise and input stage noise shaped by the National Acoustics Laboratory-Revised (NAL-R) response.
Main Results:
- The perceived audible and objectionable noise levels were recorded for each subject.
- Overall noise levels and spectra were analyzed for both noise types.
- A correlation was found between 1/3-octave noise measurements, audiometric configuration, and the audibility of internal noise.
Conclusions:
- Internal amplifier noise audibility can be predicted based on audiometric configuration and 1/3-octave noise measurements.
- Simulated input-stage and output-stage noise yielded similar audibility thresholds.
- Findings support improved hearing aid fitting by considering internal noise perception.