Changes in platelet, granulocyte, and complement activation during cardiopulmonary bypass using heparin-coated equipment
Related Experiment Video
Updated: Nov 4, 2026

Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
M Fukutomi1, S Kobayashi, K Niwaya
1Department of Surgery III, Nara Medical College, Japan.
Insights
Heparin-coated cardiopulmonary bypass (CPB) systems significantly reduce platelet, granulocyte, and complement activation. This biocompatible approach also lowers postoperative blood loss in patients undergoing cardiac surgery.
Area of Science:
- Biomedical Engineering
- Cardiovascular Surgery
- Materials Science
Background:
- Cardiopulmonary bypass (CPB) systems can trigger inflammatory responses, affecting platelets, granulocytes, and complement activation.
- Improving the biocompatibility of CPB circuits is crucial for reducing complications during cardiac surgery.
Purpose of the Study:
- To investigate the impact of heparin-coated CPB systems on inflammatory markers and clinical outcomes.
- To compare the effects of a fully heparin-coated CPB system versus a partially coated system and an uncoated system.
Main Methods:
- A comparative study involving 30 patients undergoing coronary artery bypass surgery.
- Three groups were analyzed: fully heparin-coated CPB (HC), heparin-coated oxygenator with uncoated circuit (HO), and uncoated CPB (UC).
- Measurements included plasma C3a, C4a, granulocyte elastase, beta-thromboglobulin, platelet counts, and postoperative blood loss.
Main Results:
- The heparin-coated CPB system (HC) significantly reduced complement activation (C3a, C4a) and granulocyte activation (elastase) compared to uncoated systems.
- The HC group showed significantly higher platelet counts and lower beta-thromboglobulin levels, indicating reduced platelet activation.
- Postoperative blood loss was significantly lower in the HC group compared to the uncoated group.
Conclusions:
- Heparin coating of the entire CPB circuit enhances biocompatibility by mitigating inflammatory responses.
- The HC CPB system effectively reduces platelet, granulocyte, and complement activation.
- This improved biocompatibility translates to reduced postoperative bleeding, highlighting the clinical benefits of heparin-coated CPB systems.
Abstract:
The effects of heparin-coated cardiopulmonary bypass (CPB) systems on platelet, granulocyte, and complement activation were investigated during cardiopulmonary bypass. Thirty patients underwent coronary artery bypass surgery with a heparin-coated (Carmeda Bio-Active Surface, CBAS, Medtronic, U.S.A.) CPB system (HC group, n = 10), a heparin-coated oxygenator and uncoated CPB circuit (HO group, n = 10), or an uncoated system (UC group, n = 10). In the HO group, plasma C3a (1667 +/- 632 ng/ml) and C4a (1088 +/- 319 ng/ml) concentrations were significantly (p < 0.05) lower than in the UC group (2846 +/- 1045 ng/ml and 1494 +/- 480 ng/ml, respectively) 10 min after the administration of protamine, but there were no significant differences in the platelet or granulocyte counts. In the HC group, granulocyte elastase concentrations 120 min after the onset of CPB (365 +/- 177 micrograms/L) and 10 min after the administration of protamine (676 +/- 314 micrograms/L) were significantly (p < 0.05) lower than in the other 2 groups (820 +/- 341 and 893 +/- 303 micrograms/L and 1365 +/- 595 and 1,258 +/- 622 micrograms/L). In addition, the increase in the plasma C3a concentration in the HC group 60 (p < 0.05) and 120 min after the onset of CPB (p < 0.05) was significantly less than in the other 2 groups. The C3a and C4a concentrations 10 min after the administration of protamine were significantly (p < 0.005 and p < 0.05) less in the HC group than in the UC group. Platelet counts 10 min after the administration of protamine were significantly higher (p < 0.05) and plasma beta-thromboglobulin concentrations during CPB were significantly lower in the HC group than in the other 2 groups 5 (p < 0.05), 60, and 120 min (p < 0.005) after the onset of CPB. Postoperative blood loss during the first 12 h in the HC group was significantly (p < 0.05) less than that in the UC group. The heparin-coated oxygenator and uncoated CPB circuit reduced complement activation but demonstrated no significant effects on the platelet and granulocyte systems. However, the heparin-coated CPB circuit (with all components making blood contact) reduced platelet, granulocyte, and complement activation and significantly reduced postoperative blood loss. Therefore, heparin coating of CPB systems improves biocompatibility.
More Related Videos
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
04:37Comprehensive Analysis of Procoagulant Platelets Exhibiting Features of Necrosis, Apoptosis and Platelet Activation
Published on: May 23, 2025
Related Concept Videos
Introduction to Hemostasis
The three phases of hemostasis involve many clotting factors present in plasma and several substances released by platelets and injured tissue cells. It is a fast, localized, and...
Extrinsic and Intrinsic Pathways of Hemostasis
The Extrinsic Pathway
The extrinsic pathway of coagulation is typically initiated by tissue damage that exposes blood to tissue factor (TF), a protein released by the damaged tissue cells outside the blood vessels—this interaction with TF triggers biochemical reactions involving specific clotting factors. The key player here is Factor VII, which forms a...
Anticoagulant Drugs: Low-Molecular-Weight Heparins