Five oxygen-nitrous oxide proportioning systems compared

P Feigenwinter1, T H Pohle, A M Zbinden

  • 1Institute of Anaesthesiology and Intensive Care, University Hospital of Bern, Switzerland.

European Journal of Anaesthesiology
|January 1, 1996
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Clinical validation of electromyography and acceleromyography as sensors for muscle relaxation.

European journal of anaesthesiology·2007
Same author

Model-based control of neuromuscular block using mivacurium: design and clinical verification.

European journal of anaesthesiology·2006
Same author

Model-based control of mechanical ventilation: design and clinical validation.

British journal of anaesthesia·2004
Same author

Different benefit of bispectal index (BIS) in desflurane and propofol anesthesia.

Acta anaesthesiologica Scandinavica·2003
Same author

Prediction of the haemodynamic response to tracheal intubation: comparison of laser-Doppler skin vasomotor reflex and pulse wave reflex.

British journal of anaesthesia·2002
Same author

Central hypersensitivity in chronic pain after whiplash injury.

The Clinical journal of pain·2002
Same journal

2026 ESAIC Consensus Document on Mitigation Strategies in Intensive Care Medicine: Consensus document of the European Society of Anaesthesiology and Intensive Care.

European journal of anaesthesiology·2026
Same journal

Does percutaneous cryoneurolysis spare myelinated fibres? Neurophysiological evidence after intercostal nerves freezing for pectus excavatum surgery: A retrospective pilot study.

European journal of anaesthesiology·2026
Same journal

Diminished rest-activity rhythm is associated with postoperative complications and mortality: A prospective cohort study of UK Biobank participants.

European journal of anaesthesiology·2026
Same journal

Beyond mathematical models: Why vial sizes and water safety matter.

European journal of anaesthesiology·2026
Same journal

Femoral nerve neurolysis in frail older adults with hip fracture: aligning peri-operative care with palliative goals.

European journal of anaesthesiology·2026
Same journal

2025 ESAIC and ESPA Guidelines on neuromuscular block in children.

European journal of anaesthesiology·2026
See all related articles

Contemporary anesthetic gas delivery systems effectively prevent hypoxic mixtures, even under abnormal conditions. Most systems are suitable for low-flow anesthesia, with some also supporting minimal-flow techniques.

Area of Science:

  • Anesthesiology
  • Medical Device Engineering

Background:

  • Anesthetic workstations commonly use oxygen-nitrous oxide proportioning systems to prevent hypoxic gas mixtures.
  • Ensuring accurate gas delivery is critical for patient safety during anesthesia.

Purpose of the Study:

  • To evaluate the safety and efficacy of oxygen-nitrous oxide proportioning systems in preventing hypoxic mixtures.
  • To assess the performance of different gas delivery modules across a range of fresh gas flows.

Main Methods:

  • Investigated five gas delivery modules of two types.
  • Tested fresh gas flows from 50 mL min-1 to 20 L min-1.
  • Measured oxygen concentrations at the fresh gas outlet under normal and abnormal conditions.

Main Results:

Related Experiment Videos

  • All modules successfully prevented hypoxic gas mixtures.
  • Increased oxygen concentrations were observed below 1 L min-1 fresh gas flow.
  • Most systems are suitable for low-flow anesthesia; one exception noted for minimal-flow techniques.

Conclusions:

  • The tested gas delivery modules reliably prevent hypoxic mixtures.
  • These systems are generally suitable for low-flow and minimal-flow anesthesia techniques.
  • Device performance is consistent under varying gas supply pressures.