Initial experience with 1.5-mm2 high impedance, steroid-eluting pacing electrodes
R Fogel1, F Pirzada, D Casavant
1Department of Cardiology, University Hospital, Boston, Massachusetts 02118, USA.
Pacing and Clinical Electrophysiology : PACE
|February 1, 1996
Summary
This study shows that small, steroid-eluting pacing electrodes provide effective pacing and sensing. These electrodes offer higher impedance and may improve pacemaker longevity, despite rare complications.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Permanent pacemaker implantation relies on effective lead systems.
- Smaller electrode surface areas are being explored for improved pacing efficiency and device longevity.
- Steroid-eluting electrodes aim to reduce pacing thresholds and inflammation.
Purpose of the Study:
- To evaluate the performance and safety of 1.5 mm2 unipolar steroid-eluting pacing electrodes.
- To assess pacing impedance, thresholds, and sensing capabilities over a 52-week period.
- To identify any complications associated with these novel electrode designs.
Main Methods:
- Implantation of 1.5 mm2 unipolar steroid-eluting electrodes in 64 patients (21 atrial, 62 ventricular).
- Measurement of pacing impedance, thresholds, and sensing via pacemaker telemetry and pacing systems analyzer.
- Systematic follow-up for 52 weeks to monitor lead performance and complications.
Main Results:
- Acute pacing impedances were high (1,039 ± 292 Ω atrial, 1,268 ± 313 Ω ventricular).
- A slight impedance decline occurred at 1 week, followed by stability; pacing thresholds and sensing amplitudes remained stable.
- A 6% complication rate was observed with ventricular electrodes, including dislodgement and perforation; late exit block occurred in two patients.
Conclusions:
- 1.5 mm2 steroid-eluting electrodes demonstrate excellent pacing and sensing performance with high impedances.
- These electrodes offer potential for enhanced pacing efficiency and pacemaker longevity.
- Further investigation is needed for late exit block occurrences, but the design shows promise.