Related Experiment Video
Updated: Aug 12, 2026

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
Published on: December 9, 2010
High speed diffusion imaging in the presence of eddy currents
1Department of Radiological Sciences, UCLA School of Medicine, UCLA Medical Center 90095-1721, USA.
Abstract:
This paper describes two schemes for fast diffusion imaging on a 1.5T MRI unit with unshielded gradients and 10-mT/m gradient system. The prepared longitudinal magnetization is sampled by a turbo fast low-angle shot (FLASH) sequence. The first scheme involves modifications to a diffusion-weighted driven-equilibrium preparation sequence that yields quantitative diffusion imaging. The second scheme consists of four 90 degree pulses that form the preparatory period with diffusion gradients applied between the first and last two 90 degrees pulses. Sufficient time is allowed between the diffusion gradients and the beginning of the echo readout to allow eddy currents to decay to an acceptable level. Phantom studies show agreement between diffusion coefficients in the literature with those determined using the above schemes. The potential application of these sequences in monitoring temperature is demonstrated in phantoms. Artifact-free quantitative diffusion-weighted images of the brain using the second scheme are presented.
Related Concept Videos
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Magnetic Resonance Imaging