Related Experiment Video
Updated: Aug 12, 2026

A 96 Well Microtiter Plate-based Method for Monitoring Formation and Antifungal Susceptibility Testing of Candida albicans Biofilms
Published on: October 21, 2010
Rapid determination of amphotericin B in serum and urine by third-order derivative spectrophotometry
N A Botsoglou1, D J Fletouris, G E Papageorgiou
1Laboratory of Nutrition, Aristotle University, Thessaloniki, Greece.
Abstract:
A derivative spectrophotometric method for rapid monitoring of amphotericin B in serum and urine down to 30 ng/mliters is described. Samples are treated with acetonitrile, and amphotericin B is directly quantified in the crude extracts on the basis of the intensity of the peak that appears at 402 nm when the normal absorption spectrum is submitted to third-order derivative processing. Accuracy data suggested recoveries in the range of 84.3-94.9% for serum and 85.6-93.4% for urine. The precision of the method was better than 11.3% for serum and 9.2% for urine when samples contained as low as 29.6 ng/mliters of amphotericin B. Ease of applicability, short analysis time, low cost, and reliability are the main advantages of the method.
More Related Videos
Related Concept Videos
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Antifungal Agents