Related Experiment Video
Updated: Jul 24, 2026

Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Effect of measurement error on calculated variables of oxygen transport
R P Woda1, R D Dzwonczyk, J P Orlowski
1Department of Anesthesiology, Ohio State University, Columbus 43210, USA.
Abstract:
The effect of measurement error in pH, PCO2, and PO2 on mathematically derived variables of oxygen transport in patients was delineated by comparing calculated oxygen saturations from a blood-gas machine with measured saturations from a CO-oximeter and further by modeling the error in a computer simulation. Twenty-one critically ill patients aged 30-84 yr were studied. A total of 80 arterial and 80 mixed venous blood gas samples were collected. The intraclass correlation results between measured and calculated arterial (SaO2) and mixed venous (Sv-O2) oxygen saturations were 0.59 and 0.68, respectively. The product-moment correlation for SaO2 was 0.75 and for Sv-O2 was 0.77. The percent error in calculating and measuring oxygen saturation was found to be greater at low PO2 values, whereas percent error of calculating oxygen consumption increased as the PO2 increased. Measurement repeatability at high PO2 is better than at low PO2 for both measured and calculated methods. We conclude from this comparison that measured and calculated SaO2 and Sv-O2 values are not interchangeable. Each can introduce substantial error in calculating oxygen consumption through error propagation and error amplification.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Factors Affecting Respiration
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Propagation of Uncertainty from Systematic Error
Oxygen Transport in the Blood