Lipid interference in the determination of the concentration of haemoglobin in plasma using the ACA SX analyzer

C Heller1, R Hinzmann, M Hofmann

  • 1Abteilung Klinische Chemie, Zentrallabor, Zentrallabor, Zentrum Innere Medizin, Georg-August-Universität Göttingen, Germany.

European Journal of Clinical Chemistry and Clinical Biochemistry : Journal of the Forum of European Clinical Chemistry Societies
|October 1, 1996
PubMed

In comparison to a triple wavelength procedure, the dual wavelength method for the determination of plasma haemoglobin concentration using the ACA analyzer showed considerable interference with hypertriglyceridaemic (triacylglycerols > 2.3 mmol/l) plasma. By addition of isolated human lipoprotein fractions to normotriglyceridaemic plasma, chylomicrons were identified as a major source of interference with the ACA plasma haemoglobin method, whereas VLDL was without effect up to a triacylglycerol concentration of 5.7 mmol/l. Airfuge ultracentrifugation proved to be a reliable means for removal of interfering lipid. We conclude that the extent of lipid interference with the ACA plasma haemoglobin method is highly dependent on the type of lipoprotein present. An accurate measurement of plasma haemoglobin concentrations in non-fasting plasma can only be ensured after lipid removal through airfuge ultracentrifugation.