[Comparative study of the efficiency of bacterial filters in long-term mechanical ventilation]
H J Nielsen1, P Mecke, S Tichy
1Klinik für Anästhesiologie, Medizinische Universität zu Lübeck.
Unlabelled:
Two commercially available bacterial filters to be used as part of the mechanical ventilation unit during anaesthesia were tested for hygienic criteria. Manufacturers claim that bacterial breathing filters have a filtration capacity of about 99.995%, so that there would be no need for thermal disinfection of tubing and ventilation circuits after each use. One filter is designed for a single use only, the other can be used up to 24 times after sterilisation. Both filters consist of hydrophobic glass fibres.
Methods:
During simulated mechanical ventilation for 24 h, an alcoholic suspension of Bacillus subtilis was atomized in front of the filters tested. A gelatin membrane filter was integrated in the ventilation circuit and captured the filtered gas behind the test filter.
Results:
During simulated mechanical ventilation for 24 h, the filtration capacity of both the disposable and reusable filters (Table 2) did not confirm the manufacturers' short-term technical findings over 8 s (DIN-EN 143).
Conclusions:
The use of bacterial filters during mechanical ventilation reduces the probability of bacterial contamination, but does not make sterilisation of the tubes and ventilation circuit unnecessary.