Related Experiment Video
Updated: Aug 1, 2026

Making MR Imaging Child's Play - Pediatric Neuroimaging Protocol, Guidelines and Procedure
Published on: July 30, 2009
Characterization and prediction of gradient acoustic noise in MR imagers
1GE Corporate Research and Development, Schenectady, New York 12301, USA.
Abstract:
Gradient acoustic noise has been measured and characterized for an epoxy-potted, shielded gradient assembly in a 1.5 T MRI system. Noise levels vary by 10 dB or more as a function of longitudinal position in the scanner and reflect the pattern of forces applied to the gradient assembly. The noise level increases slightly (1-3 dB) with a patient in the scanner. The spectrum of the noise is similar (but not identical) to the spectrum of the input signal. A gradient-pulse-to-acoustic-noise transfer function was obtained by using a white noise voltage input to the gradient system. The transfer function enabled us to accurately predict acoustic noise output for a pulse sequence consisting of a series of trapezoidal pulses on a single axis and for a clinical fast spin echo sequence with gradients present on all three axes.
More Related Videos
11:27Preparation and In Vitro Characterization of Dendrimer-based Contrast Agents for Magnetic Resonance Imaging
Published on: December 4, 2016
09:55Troubleshooting and Quality Assurance in Hyperpolarized Xenon Magnetic Resonance Imaging: Tools for High-Quality Image Acquisition
Published on: January 5, 2024
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI